Introduction
Intraoral scanner is a handheld optical medical device used to capture a digital impression of teeth, gingiva (gums), and other intraoral structures. Instead of using traditional impression materials (such as alginate or silicone) to create a physical mold, it creates a three-dimensional (3D) dataset that can be viewed immediately, stored, and shared for treatment planning and fabrication workflows.
In hospitals and clinics, Intraoral scanner matters because it supports modern digital dentistry and oral–maxillofacial workflows: restorative dentistry (crowns, bridges), orthodontics (aligners, retainers), implant prosthodontics, and digital documentation. It can improve operational consistency, reduce material handling, and streamline communication between chairside teams and dental laboratories—while also introducing new requirements for infection control, training, IT (information technology) support, cybersecurity, and maintenance.
This article explains what Intraoral scanner is, when it is appropriate to use, basic operation, patient safety practices, how to interpret outputs, what to do when problems occur, infection prevention and cleaning principles, and a practical global market overview for administrators and procurement teams.
What is Intraoral scanner and why do we use it?
Definition and purpose
Intraoral scanner is a digital impression system that captures the surface geometry of intraoral structures and converts it into a 3D model. The primary purpose is to create a clinically usable digital representation that can be used to:
- Design restorations and appliances using CAD/CAM (computer-aided design/computer-aided manufacturing)
- Communicate with dental laboratories and in-house production units (milling or 3D printing)
- Document baseline anatomy and track changes over time (where clinically appropriate and within local scope)
In most workflows, Intraoral scanner is used as a data acquisition step—similar in concept to taking an “optical impression”—before downstream design and manufacturing.
Common clinical settings
You will encounter Intraoral scanner across a range of settings, including:
- General dentistry clinics (private and public)
- Prosthodontics and restorative dentistry services
- Orthodontic clinics (especially for clear aligner and retainer workflows)
- Implant dentistry and prosthetic planning clinics
- Oral and maxillofacial surgery services (for prosthetic and reconstructive planning where surface scans are helpful)
- Academic dental schools and teaching hospitals
- Dental laboratories that support chairside or centralized scanning workflows
In hospital environments, Intraoral scanner may be located in dental departments, specialty clinics, operating-room-adjacent dental suites (varies by facility), or centralized digital dentistry units.
Key benefits in patient care and workflow
Benefits depend on clinical indication, operator technique, and system ecosystem, but common advantages include:
- Immediate visualization and quality feedback: teams can identify missing areas and rescan during the same visit.
- Digital storage and reproducibility: files can be archived, duplicated, and shared without physical model storage.
- Workflow efficiency: can reduce steps related to impression material mixing, tray selection, disinfection of impressions, and shipping logistics.
- Patient experience: many patients prefer optical scanning over impression trays, particularly those sensitive to taste/texture or gag reflex (tolerance varies).
- Standardized communication: digital files can support clearer lab prescriptions and iterative communication.
From an operations perspective, Intraoral scanner can shift costs from consumable impression materials toward capital equipment, software licensing, service contracts, and scanner tips/sleeves (varies by manufacturer).
How it functions (plain-language mechanism)
Most Intraoral scanner systems work by projecting a controlled light pattern (or using specialized optics) and capturing multiple images or frames as the clinician moves the scanning tip across teeth and soft tissue. Software then:
- Detects features across frames
- Aligns frames (“stitches” them) into a continuous 3D surface
- Outputs a 3D mesh model and associated textures (color may be available, varies by manufacturer)
Common underlying approaches include structured light and other optical ranging methods; the exact technology and performance characteristics vary by manufacturer and model. Some systems require powdering reflective surfaces; many newer systems are designed to be “powder-light” or “powderless,” but workflow requirements vary by manufacturer.
How medical students encounter it in training
Medical and dental trainees typically meet Intraoral scanner during:
- Preclinical simulation labs (typodont scanning, learning scan paths and common artifacts)
- Restorative dentistry and prosthodontics rotations (digital impressions for indirect restorations)
- Orthodontic clinics (aligner records and monitoring)
- Interprofessional settings (collaboration with dental technologists and biomedical engineering for device readiness)
For learners, Intraoral scanner is a practical entry point into digital health concepts: data integrity, file formats, interoperability, infection prevention, and safe use of clinical device software in patient care environments.
When should I use Intraoral scanner (and when should I not)?
Appropriate use cases
Intraoral scanner is commonly used for:
- Indirect restorations: crowns, inlays, onlays, veneers (case selection and margin visibility are critical)
- Fixed prosthodontics: short-span bridges in many workflows (full-arch complexity varies)
- Implant prosthetics: capturing implant positions using scan bodies and planning restorative steps (system compatibility varies by manufacturer)
- Orthodontics: clear aligners, retainers, study models, and progress scans
- Occlusal appliances: night guards, splints, mouthguards
- Digital documentation: baseline models for communication, teaching, and follow-up comparisons (where appropriate)
Hospitals may also use Intraoral scanner to support multidisciplinary planning for oral rehabilitation, maxillofacial prosthetics, and complex restorative cases, especially when coordination with laboratories and surgical services is needed.
Situations where it may not be suitable
Intraoral scanner may be challenging or less suitable in scenarios such as:
- Poor moisture control: heavy saliva, blood, or crevicular fluid can reduce scan quality and cause artifacts.
- Subgingival margins or deep preparation finish lines: optical scanning requires line-of-sight; margin capture can be difficult if tissues obscure the area.
- Limited access: restricted mouth opening, severe trismus, or extreme posterior access limitations.
- Significant patient movement or intolerance: scanning requires steady positioning; brief pauses can help, but tolerance varies.
- Highly reflective or complex surfaces: some materials and shiny surfaces can reduce tracking unless workflow adjustments are made (varies by manufacturer).
- Edentulous or long-span full-arch cases: full-arch accuracy and stitching stability vary by model and operator technique; conventional workflows may still be used in some settings.
Operationally, scanning may be a poor fit if the facility lacks a reliable service ecosystem (repairs, spare tips, software updates), adequate IT support, or an aligned laboratory partner.
Safety cautions and contraindications (general, non-clinical)
General safety cautions include:
- Infection prevention: the scanning tip is a semi-critical item (contacts mucous membranes). Reprocessing requirements vary by manufacturer and must match your infection prevention policy.
- Device integrity: do not use if the tip is cracked, optics are damaged, or the handpiece housing is compromised.
- Optical source precautions: avoid directing the scanning light toward eyes; follow labeling and manufacturer guidance (laser/LED class and requirements vary by manufacturer).
- Choking/aspiration risk: scanning tips and detachable parts must be secure; maintain control of components and use appropriate suction and positioning.
- Electrical and trip hazards: cables and carts can create hazards in tight operatories; manage cords and maintain dry surfaces.
Emphasize clinical judgment and supervision
Intraoral scanner supports clinical workflows but does not replace clinical judgment. Use should be supervised according to training level, local scope of practice, and institutional protocols. If scan quality is insufficient for the intended purpose, teams should escalate to an alternative workflow (which may include conventional impressions or additional documentation), consistent with local standards.
What do I need before starting?
Required setup, environment, and accessories
A typical Intraoral scanner setup includes:
- Scanner handpiece and docking/charging station (or wired connection)
- Removable scanning tips (sterilizable or single-use; varies by manufacturer)
- Computer/workstation or cart with adequate graphics performance (requirements vary by manufacturer)
- Licensed scanning software and user accounts
- Network connectivity (local network and/or internet if cloud features are used; varies by manufacturer)
- Accessories for clinical field management: high-volume suction, saliva ejector, retractors, cotton rolls, air/water syringe, mirrors
- Consumables: barrier sleeves, disinfectant wipes compatible with the device surfaces, scanner tip packaging materials for reprocessing
- Optional calibration tool or calibration fixture (if required by the specific model)
Environmental prerequisites often overlooked in procurement include stable power, adequate counter space for “clean” and “dirty” zones, and a reprocessing pathway for tips that aligns with your infection prevention program.
Training and competency expectations
Safe and effective use requires competency in:
- Patient positioning and intraoral ergonomics
- Moisture control and soft tissue management
- Scanning strategy (systematic coverage to maintain tracking)
- Recognizing and correcting artifacts (holes, double surfaces, distortions)
- File handling and correct patient identification
- Cleaning and reprocessing steps (and documentation)
Training typically combines manufacturer onboarding with supervised clinical practice. Facilities often benefit from a competency checklist and periodic revalidation, especially where staff rotation is frequent (teaching hospitals, large group practices).
Pre-use checks and documentation
A practical pre-use checklist includes:
- Confirm the device is assigned to the correct clinical area and has passed required safety checks (per facility policy).
- Inspect the handpiece and cable for damage; confirm the tip seats securely.
- Confirm the tip is appropriately reprocessed and packaged for point-of-care use (method varies by manufacturer).
- Ensure software is functioning and the correct user is logged in (audit trails matter).
- Verify patient identity in the software and match to the clinical record to prevent data mis-association.
- Check storage location settings (local vs cloud) and confirm the intended lab destination.
Documentation practices vary by jurisdiction and organization. At minimum, record that a digital impression was taken, the intended purpose, and where it was sent/stored per policy.
Operational prerequisites: commissioning, maintenance readiness, consumables, and policies
From an operations and biomedical engineering perspective, “ready to use” usually means:
- Commissioning/acceptance testing completed per facility process (functional check, baseline scan verification, asset tagging).
- A preventive maintenance plan exists (frequency varies by manufacturer and risk assessment).
- Reprocessing workflow is validated with infection prevention and central sterilization (if applicable).
- Consumables are stocked (tips, sleeves, approved wipes) with clear reorder thresholds.
- IT and cybersecurity requirements are met (patching responsibilities, user access controls, backups, and incident response pathways).
- Policies exist for data retention, export formats, and lab communication.
Roles and responsibilities (clinician vs biomedical engineering vs procurement)
Clear ownership prevents downtime and safety gaps:
| Function | Typical owner | Examples of responsibilities |
|---|---|---|
| Clinical use | Clinician/dentist and chairside team | Case selection, scanning technique, verification, patient communication, documentation |
| Infection prevention | Infection control team + reprocessing unit | Approved disinfectants, tip sterilization method, workflow audits, staff training |
| Biomedical engineering (clinical engineering) | Biomedical engineering | Asset management, safety checks, service coordination, repair triage, loaner management |
| IT / cybersecurity | IT department | Workstation build, accounts/permissions, network access, updates, data security, backups |
| Procurement | Procurement / supply chain | Vendor selection, contract terms, consumables sourcing, total cost of ownership evaluation |
| Dental laboratory interface | Lab manager / clinical lead | File acceptance criteria, turnaround workflow, remake feedback loop |
In many organizations, success depends less on the scanner itself and more on the surrounding system: training, reprocessing, IT reliability, and lab integration.
How do I use it correctly (basic operation)?
Workflows vary by model and software, but the following steps are commonly universal.
1) Prepare the patient and clinical field
- Explain what the device does and what the patient will feel (gentle contact, light, suction).
- Position the patient for operator ergonomics and visibility.
- Achieve moisture control with suction and isolation as appropriate.
- Retract soft tissues to expose margins and interproximal areas.
A consistent field (dry, well-retracted) is one of the strongest predictors of scan quality across systems.
2) Prepare the device and software
- Confirm the correct scanning tip is in place and ready for use (sterile/clean status per your protocol).
- Launch the scanning software and select the correct patient record.
- Select the scan type: single tooth, quadrant, full arch, or bite registration (options vary by manufacturer).
- Calibrate if required by the device or if prompted by the software (calibration frequency varies by manufacturer).
3) Start scanning with a systematic path
A common approach is:
- Begin on occlusal (biting) surfaces to establish tracking.
- Extend to lingual/palatal surfaces, then buccal surfaces.
- Keep a steady speed and consistent distance; avoid abrupt jumps that can cause loss of tracking.
- Capture preparation margins carefully by angling the tip and using retraction.
If tracking is lost, most systems allow you to return to a previously captured, feature-rich area and resume.
4) Manage problem areas in real time
- Pause to dry pooled saliva; rescan rather than accept a compromised region.
- Address fogging by following the manufacturer’s approach (some tips have heating elements; others rely on technique).
- For reflective surfaces or metal, adjust angle and scanning strategy; the need for powdering varies by manufacturer.
5) Verify completeness and quality
Before finalizing:
- Rotate the 3D model and check for holes, distortions, or double surfaces.
- Confirm margins and interproximal contacts are clearly captured for the intended procedure.
- Confirm that the scan includes sufficient adjacent anatomy for seating and occlusal relationships.
6) Capture occlusion (bite registration) if needed
- Acquire bite scans according to software instructions.
- Verify that the maxillary and mandibular scans articulate correctly.
- Be cautious: bite scan errors can propagate into restoration design issues even if individual arch scans look good.
7) Finalize, export, and communicate
- Finalize the scan and save with correct naming conventions.
- Export in the required file format (commonly STL for geometry; PLY/OBJ may include color—varies by manufacturer and settings).
- Send to the lab or downstream CAD/CAM system using the approved pathway.
- Document the scan and lab prescription in the clinical record.
Typical settings and what they generally mean (non-brand-specific)
Depending on the system, you may see settings such as:
- Resolution/quality mode: higher detail may increase scan time and file size.
- Color on/off: color texture can aid visualization but may not be required for all manufacturing workflows.
- Scan region selection: limiting to quadrant or preparation area can reduce file size and stitching complexity.
- Export mode: “open” exports (e.g., STL) versus proprietary formats; availability varies by manufacturer.
For administrators, “workflow lock-in” often depends on export formats, subscription requirements, and lab ecosystem compatibility—details that should be confirmed during procurement.
How do I keep the patient safe?
Patient safety with Intraoral scanner is primarily about infection prevention, device integrity, soft tissue protection, and human factors.
Core safety practices at the chairside
- Confirm identity and correct record: mismatching scans to the wrong patient is a high-impact, preventable error.
- Use appropriate retraction and suction: helps prevent aspiration risk and improves visibility.
- Avoid tissue trauma: gentle contact; do not lever against teeth or gingiva with the tip.
- Use breaks when needed: long scans can strain jaw opening; brief pauses can improve tolerance.
- Maintain situational awareness: cables, carts, and foot traffic can create trip hazards in busy clinics.
Optical and thermal considerations
- Many scanners use bright light sources, and some incorporate heating for anti-fog. Follow the manufacturer’s labeling and instructions.
- Do not intentionally point the scanning light toward eyes; keep scanning within the mouth as designed.
- If the patient reports unexpected heat, sharp discomfort, or you notice tip overheating, stop and assess; causes and thresholds vary by manufacturer.
Electrical and mechanical safety
- Inspect cables and connectors; do not use damaged equipment.
- Keep liquids away from electrical connections and docking stations.
- Ensure the tip locks correctly and cannot detach easily during use.
Human factors and team behaviors
- Standardize a “scan timeout” similar to other safety checks: correct patient, correct procedure, correct arch, correct lab destination.
- Use a clean/dirty workflow: avoid touching keyboards or phones with contaminated gloves; assign roles if possible.
- Foster an incident reporting culture: near-misses (wrong patient selected, contaminated tip discovered) should be reported and used for process improvement, not blame.
Follow facility protocols and manufacturer guidance
Your facility’s policies and the manufacturer’s IFU (instructions for use) should define:
- Approved disinfectants and sterilization parameters
- Tip lifecycle limits and inspection criteria
- Required calibration and maintenance intervals
- Acceptable software and cybersecurity practices
If policies conflict, escalate through infection prevention, biomedical engineering, and clinical governance rather than improvising at the chairside.
How do I interpret the output?
Types of outputs/readings
Intraoral scanner typically produces:
- A 3D surface mesh of the scanned area (digital model)
- Color texture mapping (if enabled and supported)
- Alignment of upper and lower arches with bite registration
- Measurement tools (distance, arch width, tooth size) within the software environment
- Annotation tools (margin marking, notes to lab)
These outputs are most often used for fabrication workflows rather than direct diagnosis.
How clinicians typically interpret the scan
Clinicians and lab teams generally assess:
- Completeness: no missing regions critical to seating or margin integrity
- Surface fidelity: no obvious distortions, “ripples,” or duplicated surfaces
- Margin clarity: finish lines and emergence profiles are visible where needed
- Interproximal capture: contact areas are adequately represented for design
- Occlusal relationship: bite alignment makes anatomical sense and matches clinical observation
A practical approach is to interpret the scan as a data capture step: the question is whether the dataset is adequate for the intended downstream task.
Common pitfalls and limitations
Common limitations include:
- Artifacts from saliva, blood, or fogging: can create holes or warped surfaces.
- Soft tissue movement: tongue and cheeks can be mistakenly captured, confusing the stitching process.
- Stitching drift on long scans: full-arch accuracy can be more sensitive to technique and system capability.
- Subgingival data gaps: optical systems cannot capture what they cannot “see.”
- Bite registration errors: small misalignments can cause major occlusal design problems.
Clinical correlation is essential
A scan that “looks good” on screen can still be clinically inappropriate if:
- margins were not adequately exposed,
- the preparation is flawed,
- the bite record is not representative, or
- the clinical plan has changed.
Use the scan in conjunction with clinical examination and any other required documentation per your local standards. Intraoral scanner output should be interpreted within the clinical context and the intended use of the dataset.
What if something goes wrong?
Troubleshooting checklist (practical and non-brand-specific)
If scanning is not working as expected, consider:
Scan quality problems
- Confirm the tip is clean and not scratched or fogged.
- Re-establish dryness and retraction; rescan problem areas.
- Return to a stable, previously captured region to regain tracking.
- Check whether the software is displaying “holes” due to missed angles; adjust tip orientation.
Device and software problems
- Confirm the tip is seated correctly and recognized by the system (varies by manufacturer).
- Restart the scanning software; if needed, reboot the workstation.
- Check for low battery or unstable power connections.
- Verify network connectivity if cloud syncing or lab upload is required.
- Confirm storage space is available and the correct patient record is selected.
Workflow problems
- Confirm file format requirements for the intended lab or CAD/CAM workflow.
- Ensure the correct arch/quadrant and bite records are captured.
- Recheck naming conventions to avoid sending the wrong file.
When to stop use
Stop scanning and escalate if:
- The tip or handpiece is visibly damaged, cracked, or cannot be secured.
- The device produces unexpected heat, smell, smoke, or abnormal noise.
- There is any electrical safety concern (shock, sparking, liquid ingress).
- You cannot meet infection prevention requirements (e.g., no reprocessed tips available).
- A patient safety event occurs (soft tissue injury, aspiration risk, significant distress).
When to escalate (biomedical engineering, IT, manufacturer)
Escalation pathways often look like this:
- Biomedical engineering/clinical engineering: hardware faults, damage, power issues, docking/charging failures, scheduled maintenance, asset tracking.
- IT support: login problems, workstation performance, network connectivity, cybersecurity constraints, backups, software deployment.
- Manufacturer or authorized service provider: persistent error codes, calibration failures, replacement parts, warranty questions, software bugs (as directed by policy).
Documentation and safety reporting expectations
Good documentation supports patient safety and system improvement:
- Record what happened, when, and which patient(s) were affected (per policy).
- Capture error codes/screenshots if allowed and appropriate.
- Note device identifiers (asset tag, model, serial number) and the tip type used.
- Report incidents through your facility reporting system and follow any local regulatory reporting requirements (process varies by jurisdiction).
Infection control and cleaning of Intraoral scanner
Infection prevention is one of the most important operational determinants of safe Intraoral scanner use because the device contacts mucous membranes and is used repeatedly across patients.
Cleaning principles: what matters most
Key principles that apply across manufacturers:
- Follow the IFU: the manufacturer’s validated cleaning, disinfection, and sterilization method is the reference standard.
- Use Spaulding classification logic: items contacting mucous membranes are typically semi-critical and often require sterilization or high-level disinfection; exact requirements depend on tip design and IFU.
- Separate clean and dirty workflows: prevent cross-contamination between chairside scanning, reprocessing, and storage.
- Protect optics and electronics: many scanner handpieces must not be immersed; liquids in vents or connectors can damage the device.
Disinfection vs sterilization (general)
- Cleaning removes debris and reduces bioburden; it is usually required before any disinfection or sterilization.
- Disinfection reduces microorganisms; levels (low/intermediate/high) depend on chemical and contact time.
- Sterilization aims to eliminate all microbial life forms, typically using steam under pressure (autoclave) where compatible.
Scanner tips are often designed to be sterilizable, but not all are; some are single-use. The correct process varies by manufacturer and local policy.
High-touch points to include in your workflow
Even if only the tip enters the mouth, contamination spreads to:
- Handpiece grip area and buttons
- Cable or connector near the handpiece
- Docking station surfaces
- Workstation keyboard, mouse, touchscreen, and foot controls
- Chairside countertops and the “scan cart” handle
A common failure mode is a clean tip used with a contaminated handpiece or workstation interface.
Example cleaning and reprocessing workflow (non-brand-specific)
Your facility’s detailed steps must follow the IFU, but a typical safe pattern is:
- At point of care: remove gross debris from the tip using approved methods; avoid splashing.
- Remove the tip: detach according to manufacturer instructions; keep track of small parts.
- Contain for transport: place reusable tips in a closed, labeled container for reprocessing.
- Reprocess the tip: clean, rinse, dry, then sterilize or disinfect per IFU (cycle parameters vary by manufacturer).
- Disinfect the handpiece exterior: wipe using an approved disinfectant with correct contact time; avoid fluid entry into ports and seams.
- Protect electronics: do not immerse the handpiece unless explicitly permitted by IFU.
- Dry and inspect: check the tip’s optical window/mirror for residue, scratches, and cracks; remove from service if compromised.
- Store correctly: keep reprocessed tips in a clean, protected environment to maintain their state until use.
Operational controls for consistent infection prevention
For managers and biomedical engineering teams, reliability comes from systems:
- Maintain sufficient tip inventory to support throughput while tips are in reprocessing.
- Track tip reprocessing cycles if the manufacturer specifies a maximum number (varies by manufacturer).
- Standardize disinfectant products across the clinic to avoid material incompatibility.
- Audit real-world practice (glove-to-keyboard contamination is common).
- Coordinate with central sterile services if tips are sterilized in a centralized unit.
Infection prevention is not just a clinical issue—it is a core operational requirement for this piece of hospital equipment.
Medical Device Companies & OEMs
Manufacturer vs. OEM (Original Equipment Manufacturer)
A manufacturer is the company that markets the final medical device under its brand and is typically responsible for regulatory compliance, labeling, post-market surveillance, and customer support pathways (responsibilities vary by jurisdiction).
An OEM (Original Equipment Manufacturer) is a company that may produce components or complete hardware/software that another brand sells under its own label. In digital dentistry, OEM relationships can affect:
- Availability of spare parts and repair pathways
- Software update cadence and compatibility
- Long-term support and end-of-life planning
- Training and service coverage in different regions
For procurement, it is reasonable to ask whether a product is built in-house or relies on OEM components, and how that affects service continuity.
Top 5 World Best Medical Device Companies / Manufacturers
The following are example industry leaders (not a ranking) commonly associated with digital dentistry and/or Intraoral scanner ecosystems. Availability, model portfolios, and regional support vary by manufacturer.
-
Dentsply Sirona
A widely recognized global dental equipment company with product lines across imaging, CAD/CAM workflows, and chairside systems. In many markets, it is associated with integrated restorative workflows that combine scanning, design, and manufacturing. Global footprint and service models vary by region and dealer networks. -
Align Technology
Known globally for orthodontic systems and digital workflows that often integrate scanning with treatment planning. Intraoral scanner offerings are commonly positioned around orthodontic and restorative digital records. Support structure typically depends on country presence and authorized channels. -
3Shape
A dental technology company associated with scanning and CAD software ecosystems across clinics and laboratories. Its reputation is closely linked to digital workflows and interoperability discussions (export options and integrations vary by product and licensing). Global availability depends on distributor and partner networks. -
Planmeca
A dental equipment manufacturer with a broader portfolio that can include dental imaging, units, and digital dentistry tools depending on the market. Intraoral scanner solutions may be offered as part of integrated clinic technology stacks. Implementation quality often depends on local service capability. -
Medit
Known for Intraoral scanner products and software ecosystems in many countries, often distributed through regional partners. As with other manufacturers, device performance, features, and service coverage depend on model and location. Procurement teams typically evaluate the local support ecosystem alongside the hardware.
Vendors, Suppliers, and Distributors
Role differences: vendor vs supplier vs distributor
These terms are often used interchangeably, but operationally they can differ:
- A vendor is any entity that sells goods or services to your facility (including a manufacturer selling direct).
- A supplier is a source of products/consumables; in healthcare operations, this often emphasizes supply continuity, pricing, and contracts.
- A distributor is an intermediary that handles logistics, local inventory, installation coordination, training, and sometimes first-line service.
For capital clinical devices like Intraoral scanner, distributor capability can be as important as the scanner brand—especially for uptime, spare tips, loaner policies, and on-site training.
Top 5 World Best Vendors / Suppliers / Distributors
The following are example global distributors (not a ranking) that are commonly discussed in dental/medical supply contexts. Coverage for Intraoral scanner varies by country, authorization status, and product line.
-
Henry Schein
A large distributor with dental and medical supply operations in multiple regions. Often serves group practices, hospitals with dental departments, and academic institutions through contract-based purchasing. Service offerings and product availability vary by country and local subsidiaries. -
Patterson Dental (Patterson Companies)
A major dental distributor primarily associated with North American markets. Typically supports clinics with equipment sales, practice integration, and supply chain services. International availability and product authorization vary. -
Benco Dental
A distributor focused largely on the United States market, often providing equipment procurement support, education, and practice solutions. For buyers, value may come from implementation support and training logistics. Global reach is not uniform. -
Dental Axess
A distributor known in parts of Europe for digital dentistry equipment and workflow support (availability varies by country). Often serves clinics and laboratories seeking integrated scanning-to-production ecosystems. Local service capability depends on regional structure. -
Pluradent Group
A European dental distribution group operating through regional companies (coverage varies). Typically serves dental clinics and laboratories with equipment, consumables, and support services. Cross-border procurement and service models depend on local entities.
Global Market Snapshot by Country
India
Demand for Intraoral scanner is driven by growth in private dentistry, corporate clinic chains, and expanding interest in digital workflows for restorations and orthodontics. Many facilities rely on imported systems, making distributor support and spare-part availability important. Access is often concentrated in metro areas, with variable service reach in smaller cities.
China
China has large urban dental markets and increasing digital dentistry adoption, supported by a growing domestic medical equipment ecosystem alongside imports. Hospital and clinic buyers may evaluate both international brands and local options based on service coverage and cost. Uptake is typically higher in urban centers, with uneven access in rural regions.
United States
In the United States, Intraoral scanner is widely integrated into restorative and orthodontic workflows, supported by a mature dental laboratory network and established service infrastructure. Procurement decisions often emphasize interoperability, cybersecurity posture, and subscription/licensing transparency. Adoption is broad across private practice and institutional settings, though implementation quality varies by training and workflow design.
Indonesia
Indonesia’s market is shaped by growth in private dental services in major cities and increasing patient demand for modern restorative and orthodontic care. Systems are often imported, so reliable local distributors and training pathways are key for sustained uptime. Access can be limited outside urban areas where service and reprocessing infrastructure may be constrained.
Pakistan
In Pakistan, adoption is often led by urban private clinics and teaching institutions with interest in digital workflows. Cost sensitivity, import dependence, and variable availability of trained support personnel can affect purchasing decisions. Service coverage and spare tip logistics may differ substantially between major cities and smaller regions.
Nigeria
Nigeria’s demand is primarily concentrated in urban private clinics and larger centers with capacity for capital medical equipment investment. Import reliance can make after-sales support, power stability, and consumable availability critical considerations. Rural access is typically limited, and service ecosystems may be fragmented.
Brazil
Brazil has a large dental care sector with strong professional demand for restorative and orthodontic services, supporting interest in digital impressions. Buyers may encounter a mix of imported solutions and regional distribution capabilities. Adoption is typically stronger in major urban regions, with variability in public sector procurement and maintenance support.
Bangladesh
Bangladesh shows growing interest in digital dentistry in urban clinics and higher-volume centers, often relying on imported systems and distributor-led training. Service continuity and consumable supply can be limiting factors, particularly outside major cities. Implementation frequently depends on local technical expertise and stable IT infrastructure.
Russia
Russia’s market can involve a mix of imported and regionally sourced dental technology, with procurement influenced by regulatory pathways, currency considerations, and service availability. Larger cities tend to have stronger distribution and technical support networks. Rural adoption may be constrained by logistics and limited access to specialized training.
Mexico
Mexico’s demand is influenced by private dentistry growth and dental tourism in certain regions, which can incentivize investment in digital workflows. Many systems are imported, making authorized distribution and service contracts important for continuity. Adoption is typically higher in urban and tourism-linked areas than in rural settings.
Ethiopia
In Ethiopia, access to Intraoral scanner is generally limited to larger urban centers and specialized clinics due to capital cost, import dependence, and evolving service infrastructure. Training opportunities and reprocessing capacity can be bottlenecks. Rural access is often constrained by broader healthcare resource limitations.
Japan
Japan’s dental technology environment is relatively mature, with high expectations for quality, process discipline, and equipment reliability. Adoption of digital impressions can be supported by strong technical services and a well-developed clinical culture, though workflow choices may be influenced by local reimbursement and practice patterns. Procurement typically emphasizes long-term support and validated reprocessing processes.
Philippines
The Philippines market is often driven by private clinics and urban centers, with increasing adoption of digital workflows for restorative and orthodontic services. Import dependence is common, so distributor training and service responsiveness are central. Access outside metropolitan areas can be limited by logistics and variable technical support availability.
Egypt
In Egypt, demand is often concentrated in urban private clinics and larger institutions seeking modern restorative and orthodontic solutions. Many systems are imported, making after-sales service and consumable supply planning important. Differences between urban and rural access can be pronounced due to infrastructure and workforce distribution.
Democratic Republic of the Congo
In the Democratic Republic of the Congo, adoption is typically limited to a small number of urban centers because of cost, import logistics, and constrained service ecosystems. Facilities may face challenges with reliable power, network access, and equipment maintenance capacity. Where used, success often depends on strong distributor support and simplified workflows.
Vietnam
Vietnam has growing demand in urban areas driven by an expanding middle class, private dental chains, and interest in faster restorative workflows. Systems are frequently imported, with adoption supported by expanding training programs and distributor networks. Rural access remains variable, often limited by service reach and capital budgets.
Iran
In Iran, procurement can be influenced by import constraints and reliance on regional supply pathways, affecting brand availability and service continuity. Local technical skills may support maintenance in some centers, but access to official updates, parts, and subscriptions can vary. Urban centers typically have stronger adoption and support compared with rural areas.
Turkey
Turkey’s market is supported by a strong private dental sector and dental tourism, which can accelerate adoption of digital impression workflows. Distribution and training ecosystems are relatively active in major cities, though support quality varies by vendor. Buyers may prioritize rapid service turnaround and interoperability with diverse laboratory partners.
Germany
Germany’s market is characterized by a mature dental laboratory ecosystem and high expectations for documentation, quality systems, and regulatory compliance. Adoption is supported by strong service networks and professional training infrastructure, with procurement often attentive to data protection and lifecycle support. Access is broad, though smaller practices may adopt at different rates based on workflow fit.
Thailand
Thailand’s demand is influenced by private hospitals and clinics, including dental tourism-linked services that benefit from efficient digital workflows. Many systems are imported, so service contracts, training, and consumable supply planning are key. Adoption is typically higher in Bangkok and major centers, with variable reach in rural regions.
Key Takeaways and Practical Checklist for Intraoral scanner
- Intraoral scanner is primarily a digital impression tool that captures surface geometry for downstream workflows.
- Define the intended use (restoration, orthodontics, implant prosthetics, documentation) before selecting a scanning protocol.
- Treat the scan as clinical data: correct patient selection and file labeling prevent high-impact errors.
- Build moisture control into the workflow; saliva and blood are frequent causes of artifacts.
- Use systematic scan paths to reduce stitching errors and tracking loss.
- Verify scan completeness before finalizing; rescanning chairside is cheaper than remakes later.
- Recognize that subgingival margins may be difficult to capture with any optical system.
- Confirm whether the scanner tip is sterilizable or single-use; this varies by manufacturer.
- Keep a clean/dirty separation at the workstation to avoid contaminating keyboards and touchscreens.
- Use only disinfectants approved in the IFU to prevent material damage and premature device failure.
- Inspect tips for scratches, cracks, and residue; optical damage can degrade scan fidelity.
- Calibrate when required or prompted; calibration routines vary by manufacturer and model.
- Standardize who owns updates: IT for software, biomedical engineering for hardware, clinical leads for workflow validation.
- Plan for consumables and turnaround time in reprocessing; inadequate tip inventory causes bottlenecks.
- Confirm export formats (STL/PLY/OBJ) and lab compatibility during procurement, not after installation.
- Consider data governance early: storage location, retention, access control, and audit trails.
- Avoid assumptions about “open” vs “closed” ecosystems; licensing and features vary by manufacturer.
- Train clinicians to recognize common artifacts (holes, double surfaces, distortion) and correct them immediately.
- Capture bite registrations carefully; occlusion errors can affect restoration fit even with good arch scans.
- Stop use if the handpiece or tip is damaged or if abnormal heat, smell, or electrical issues occur.
- Escalate hardware faults to biomedical engineering and software/network faults to IT using clear pathways.
- Document device identifiers and error codes when reporting incidents to improve service turnaround.
- Include Intraoral scanner in preventive maintenance and asset management programs like other hospital equipment.
- Evaluate total cost of ownership: subscriptions, warranties, service contracts, and consumables.
- Ensure reprocessing capacity matches patient volume, especially in teaching clinics with peak loads.
- Use checklists for patient identity, arch selection, and lab destination to reduce human-factor errors.
- Validate new workflows with the dental lab to align margin capture expectations and file acceptance criteria.
- Plan for downtime with loaner options or backup workflows (digital or conventional) per local policy.
- Include infection prevention leaders in purchasing decisions to confirm reprocessing feasibility.
- Ensure electrical safety and cable management to reduce trip hazards in busy operatories.
- Do not treat scan output as a standalone diagnostic claim; correlate with clinical examination and required records.
- Provide structured onboarding for trainees, then supervised scanning until competency is demonstrated.
- Use consistent naming conventions and version control when rescanning to avoid sending outdated files.
- Confirm cybersecurity responsibilities for cloud features, remote support tools, and data export.
- Keep vendor contacts, service-level expectations, and escalation steps visible to frontline staff.
- Audit real-world compliance periodically; gaps often occur at the keyboard, cart handle, and docking station.
- Reassess scanner utilization quarterly to ensure the device delivers workflow value and patient-centered care.
- Align purchasing with the local service ecosystem; support availability can matter as much as brand selection.
- Build a culture where staff report near-misses (wrong patient selected, contaminated tip found) without blame.
- Maintain written local protocols that match the manufacturer IFU and are accessible at point of use.
If you are looking for contributions and suggestion for this content please drop an email to contact@myhospitalnow.com