Introduction
A Refrigerated centrifuge is a laboratory medical device that spins biological samples at controlled speeds while actively regulating temperature. In hospital and clinic settings, it is most often used to separate components of blood, body fluids, and cell suspensions for diagnostic testing, transfusion services, and certain therapeutic preparations. Temperature control matters because many analytes (measurable substances in a sample) and cellular components can change when warmed, especially during high-speed spinning where friction can generate heat.
For medical students and trainees, the Refrigerated centrifuge sits at the intersection of pre-analytical quality (everything that happens to a sample before it is tested) and real-world patient care. A well-run centrifugation step supports accurate results; a poorly run one can contribute to delays, sample rejection, repeat phlebotomy, or misleading laboratory values.
For hospital administrators, biomedical engineers, and procurement teams, this piece of hospital equipment is also a capacity and reliability asset: it affects laboratory throughput, biosafety risk management, staff training needs, maintenance planning, and service contract strategy.
This article explains what a Refrigerated centrifuge is, when and when not to use it, how basic operation works, how to build safe workflows, how to interpret the “output” (both the separated sample and the device parameters), what to do when problems occur, how to clean and disinfect it, and how the global market looks across diverse health systems.
What is Refrigerated centrifuge and why do we use it?
Clear definition and purpose
A Refrigerated centrifuge is a centrifuge with an integrated cooling system designed to maintain a user-set temperature during operation. Like all centrifuges, it separates particles in a liquid by spinning them rapidly so that denser components move outward and lighter components remain closer to the center. The “refrigerated” part adds controlled cooling to protect temperature-sensitive specimens and to support consistent, repeatable laboratory processing.
In healthcare, the purpose is usually one or more of the following:
- Separate blood into components (for example, to obtain plasma, serum, or a cellular layer) for laboratory testing.
- Pellet cells or microorganisms from liquid for further testing (for example, molecular workflows, certain microbiology steps, or cytology preparation).
- Concentrate or clarify samples (for example, removing debris from supernatant fluid before analysis).
This device is commonly treated as medical equipment when used in clinical diagnostic workflows, even though many models are also used in research settings. Regulatory classification and labeling requirements vary by country and intended use.
Common clinical settings
You may see Refrigerated centrifuge units in:
- Core clinical laboratories (chemistry, immunology, hematology pre-processing areas)
- Blood banks and transfusion services
- Microbiology and molecular biology laboratories
- Pathology specimen processing areas
- Fertility/IVF laboratories (highly protocol-driven; exact requirements vary by facility)
- Research laboratories within academic medical centers
- Public health laboratories and reference labs
In smaller facilities, one Refrigerated centrifuge may serve multiple benches and workflows. In larger hospitals, different models may be dedicated to specific sample types, volumes, or biosafety levels.
Key benefits in patient care and workflow
A Refrigerated centrifuge can support patient care and operations by:
- Preserving sample integrity for temperature-sensitive assays and cellular components (the degree of benefit depends on the test and protocol).
- Improving repeatability when standardized temperature is part of the method.
- Reducing workflow delays by allowing fast, controlled separation without moving samples between separate cold environments during the run.
- Supporting biosafety workflows when paired with aerosol-tight lids or sealed buckets (availability varies by manufacturer and accessory set).
A critical operational point: centrifugation is a common pre-analytical step, and pre-analytical issues are a known source of laboratory errors. A reliable centrifuge and consistent technique help reduce avoidable variation.
Plain-language mechanism of action (how it functions)
At a basic level, the device works like a very fast spinning carousel:
- Samples are placed into a rotor (the spinning part) using compatible tubes and adapters.
- The rotor spins at a programmed speed, creating centrifugal force that separates substances by density.
- During the spin, the refrigeration system removes heat to help maintain a target temperature (how tightly it holds temperature depends on ambient conditions, load, speed, and model design).
- The device stops, the lid unlocks (typically only when safe), and the separated sample can be removed for testing or further processing.
Two commonly encountered speed concepts:
- RPM: Revolutions per minute (how fast the rotor spins).
- RCF: Relative centrifugal force (the “g-force,” a multiple of Earth’s gravity).
Many laboratory protocols specify RCF because the same RPM can produce different forces with different rotor radii. Some devices display both; others require conversion. A commonly used relationship is:
- RCF depends on rotor radius and RPM (exact calculation and reference radius conventions vary by manufacturer).
If you are unsure, follow the protocol and the device manual; do not guess or “match a previous RPM” without verifying rotor type and radius.
Key components and terminology (what you’ll hear on the ward/lab)
- Rotor: Holds the samples and spins; can be fixed-angle or swinging-bucket.
- Buckets / carriers: Used in swinging-bucket rotors to hold tube racks.
- Adapters: Inserts that allow different tube sizes.
- Aerosol-tight lids or safety cups: Accessories that contain aerosols if a tube breaks (important for infection control and staff safety).
- Imbalance detection: Safety feature to stop the run when loads are uneven.
- Lid lock: Prevents opening during unsafe conditions.
- Brake setting: Controls deceleration; “soft brake” can reduce mixing of layers after separation.
How medical students typically encounter or learn this device in training
Medical students and residents often first encounter the Refrigerated centrifuge indirectly—through laboratory results—before seeing the device during:
- Clinical pathology or laboratory medicine teaching sessions
- Rotations in emergency medicine (rapid lab turnaround discussions)
- Hematology/oncology rotations (blood component handling and lab dependencies)
- Infectious disease or microbiology exposure (specimen handling and biosafety)
- Research electives (hands-on training under supervision)
A useful mental model for trainees: centrifugation is part of the diagnostic chain. If the chain is weak at any link—labeling, transport time, temperature control, balancing, or correct settings—patient-facing decisions can be affected downstream.
When should I use Refrigerated centrifuge (and when should I not)?
Appropriate use cases (common indications in healthcare workflows)
Use of a Refrigerated centrifuge is typically appropriate when one or more of the following apply:
- The sample or analyte is temperature-sensitive and the method requires controlled cooling.
- High-speed centrifugation is needed, and frictional heating could affect sample quality.
- Cell viability or function matters for downstream steps (for example, certain cell processing or research-adjacent clinical workflows).
- Standard operating procedures (SOPs) or validated methods specify refrigerated centrifugation for consistency.
- Biosafety practices require sealed accessories, and the chosen model supports compatible aerosol-tight components (availability varies by manufacturer).
Common sample types that may be processed in a Refrigerated centrifuge include blood tubes, plasma/serum aliquots, body fluids, urine (method-dependent), and cell suspensions. The exact appropriateness depends on the test, tube type, and the laboratory’s validated procedure.
When a non-refrigerated centrifuge may be suitable instead
A Refrigerated centrifuge may not be necessary when:
- The validated protocol specifies room-temperature centrifugation.
- The sample is not temperature-sensitive, and heat generation at the required speed is not expected to be clinically meaningful (this is workflow- and model-dependent).
- The facility has a dedicated non-refrigerated unit that better matches throughput needs (for example, high-volume routine serum separation with established ambient workflow).
- Budget, space, and energy constraints mean the refrigerated model should be reserved for workflows that truly need it.
In many laboratories, using refrigerated capacity for all samples is inefficient. A practical operational approach is to tier centrifugation resources: reserve the Refrigerated centrifuge for protocols that require it, and use standard centrifuges for routine processing when allowed by the method.
Situations where it may not be suitable (or where you should pause)
Do not use a Refrigerated centrifuge—or stop and seek supervision—if:
- The rotor, buckets, or adapters are damaged, corroded, or overdue for inspection/replacement per manufacturer guidance.
- You cannot confirm tube compatibility (tube type, maximum RCF, cap integrity, and chemical resistance).
- The run would exceed rated limits (maximum speed/RCF, maximum capacity, or temperature limits).
- The unit shows abnormal vibration, noise, repeated imbalance alarms, or error codes that are not resolved by basic checks.
- The device has condensation, frost, or fluid pooling in ways that suggest a refrigeration fault or a spill that was not properly decontaminated.
- A sample contains volatile, flammable, or reactive chemicals not approved for the device and accessories (compatibility varies by manufacturer and local safety policy).
Also consider whether the workflow should be performed inside a higher biosafety environment. Centrifugation can generate aerosols if a tube breaks or leaks, so local infection prevention and biosafety protocols matter.
Safety cautions and general contraindications (non-clinical)
Key cautions apply regardless of clinical context:
- Never run unbalanced loads. Imbalance increases mechanical stress and can cause rotor failure.
- Never bypass lid locks or safety interlocks. These are engineered controls to reduce injury.
- Do not open the lid until the rotor has fully stopped and the device indicates it is safe.
- Do not use damaged tubes (hairline cracks, compromised caps) or tubes not rated for the planned force.
- Use appropriate PPE (personal protective equipment) based on specimen risk (for example, gloves and eye protection; additional PPE based on local policy).
- Use aerosol-tight accessories for infectious or high-risk specimens when required by policy; not all workflows require them, but when they are indicated they are a key control.
Emphasize clinical judgment, supervision, and local protocols
For trainees: centrifuge settings are not “one-size-fits-all.” Follow your laboratory’s SOPs, tube manufacturer guidance, and supervision requirements. For leaders: make sure SOPs are accessible at point of use, and that deviations are handled through a documented quality process.
This article provides general information only. Local protocols, regulatory requirements, and manufacturer instructions should guide real-world use.
What do I need before starting?
Required setup and environment
A Refrigerated centrifuge is more demanding than a basic centrifuge because it must manage heat removal and may be more sensitive to ventilation and ambient conditions. Before starting, confirm:
- Stable placement on a level, vibration-resistant surface (bench-top or floor-standing per model).
- Adequate clearance and ventilation around intake/exhaust areas to prevent overheating and to support refrigeration performance.
- Appropriate power supply (voltage, frequency, grounding/earthing, and circuit capacity per manufacturer requirements).
- Ambient conditions within the manufacturer’s specified range (temperature and humidity affect cooling performance and condensation risk).
- Access control and ergonomics: safe lifting posture for rotors, clear space for loading/unloading, and a location that reduces traffic during operation.
In some settings, especially where power quality is variable, facilities may consider voltage stabilization or power conditioning. Whether this is appropriate depends on local infrastructure and manufacturer recommendations.
Accessories, consumables, and “hidden” requirements
A Refrigerated centrifuge is an ecosystem, not just a box. Common requirements include:
- Rotors appropriate to your workflow (fixed-angle, swinging-bucket, microplate; availability varies by model).
- Buckets/carriers and adapters for the tube sizes used in your facility.
- Tube racks compatible with buckets (for swinging-bucket designs).
- Consumables: tubes rated for the required RCF, caps, seals, and any liners used in safety cups.
- Aerosol-tight lids/safety cups when required for infectious specimens.
- Balancing tools: a balance scale and/or pre-filled balance tubes, depending on workflow.
- Temperature monitoring tools (if your quality system requires independent verification; approach varies by facility).
Procurement teams should note that accessory costs can be significant and may be budgeted separately. Standardizing tube types and rotor systems across sites can simplify training and inventory management.
Training and competency expectations
Because a Refrigerated centrifuge is both a mechanical hazard and a biosafety risk point, training should be structured. Competency typically includes:
- Understanding RCF vs RPM, and how to select the correct setting for a protocol.
- Correct rotor selection and installation.
- Correct balancing technique (mass balance, symmetry, and use of dummy tubes).
- Safe use of aerosol-tight accessories and spill procedures.
- Recognizing abnormal sounds, vibration, alarms, and error codes.
- Cleaning and decontamination steps per SOP and manufacturer IFU (Instructions for Use).
Facilities often document competency with checklists, supervised sign-off, and periodic reassessment. The details depend on the local quality management system.
Pre-use checks and documentation
A practical pre-use checklist commonly includes:
- Device status: no active service tags, no unresolved faults, and next preventive maintenance not overdue.
- Rotor condition: no visible cracks, corrosion, deformation, or damaged threads; rotor log updated if used.
- Buckets/adapters: intact, clean, correct size, and correctly seated.
- Lid seal and locking mechanism: clean, undamaged gasket; lid closes securely.
- Chamber condition: dry, clean, no debris; drain (if present) not clogged.
- Temperature readiness: unit pre-cooled if required; verify setpoint and actual reading (device display or independent check per local policy).
- SOP confirmation: correct protocol for sample type, tube type, and target RCF/time/temperature.
Documentation may include a paper logbook or an electronic log. Common entries are operator ID, rotor used, run parameters, anomalies, and cleaning confirmation. The extent of documentation varies by laboratory accreditation requirements and local governance.
Operational prerequisites: commissioning, maintenance readiness, consumables, and policies
Before first clinical use (or after relocation/major repair), many facilities perform commissioning steps such as:
- Installation qualification (installed correctly and safely)
- Operational checks (speed/temperature performance within acceptable tolerance)
- Workflow validation aligned to laboratory methods (what constitutes acceptable separation for your assays)
Terminology varies by institution (for example, IQ/OQ/PQ in some quality systems). The underlying goal is the same: confirm the device performs as intended in your environment.
Maintenance readiness considerations:
- Confirm service coverage, response times, and parts availability (varies by manufacturer and region).
- Establish a preventive maintenance schedule and rotor inspection routine.
- Keep a spare parts strategy for high-wear items (seals, adapters) where appropriate.
- Define downtime procedures, including backup centrifugation capacity and sample rerouting.
Policies that should be clear before routine use:
- Biosafety and specimen handling policy (including aerosol containment expectations)
- Cleaning/disinfection policy and approved chemicals list
- Incident reporting and escalation policy
- Access control policy (who is allowed to run which protocols)
Roles and responsibilities (clinician vs biomedical engineering vs procurement)
In a well-run hospital, responsibilities are explicit:
- Clinicians/trainees: follow SOPs, ensure correct sample identification and handling, and report deviations; they should not improvise centrifuge settings without laboratory oversight.
- Laboratory staff (medical technologists/scientists): own validated protocols, training, daily checks, quality control documentation, and first-line troubleshooting.
- Biomedical engineering/clinical engineering: installation support, preventive maintenance coordination, electrical safety, service liaison, and lifecycle management planning.
- Procurement/supply chain: vendor evaluation, tendering, contract negotiation, accessory standardization, and ensuring total cost of ownership is considered (device + rotors + service + downtime risk).
- Infection prevention and biosafety teams: define cleaning/disinfection requirements and high-risk specimen workflows.
- Facilities/engineering: power quality, ventilation/airflow, and physical space planning.
Clarifying these roles reduces “grey zone” failures where everyone assumes someone else is responsible.
How do I use it correctly (basic operation)?
Workflows vary by model and by laboratory method. The steps below reflect widely applicable practice for a Refrigerated centrifuge, but you should always follow your facility SOP and the manufacturer’s IFU.
Basic step-by-step workflow (commonly universal)
-
Confirm the protocol
Verify sample type, tube type, target RCF (or RPM), run time, temperature setpoint, and brake setting required by the SOP. -
Prepare the device
Check that the correct rotor is installed and secured according to the IFU. Ensure the chamber is clean and dry. Pre-cool the centrifuge or rotor if the workflow requires temperature stability. -
Select the correct rotor and accessories
Use only compatible buckets, adapters, and aerosol-tight lids rated for your planned RCF. “Almost fitting” adapters are a common source of tube deformation or breakage. -
Inspect tubes and cap integrity
Ensure tubes are intact, properly capped, and not overfilled. Leaks under centrifugal force can contaminate the chamber and create aerosols. -
Balance the load
Balance by mass, not by volume. Place tubes symmetrically across the rotor. Use balance tubes when needed. Even if the device has imbalance detection, do not rely on it as your primary control. -
Load samples safely
Seat tubes fully in adapters/buckets. Confirm bucket swing is free (for swinging-bucket rotors). Close and secure any safety lids. -
Close the lid and program settings
Set temperature, RCF/RPM, time, acceleration, and brake (if adjustable). Some models allow stored programs—confirm you selected the correct one. -
Start the run and monitor
Stay nearby at startup to detect immediate imbalance or abnormal noise. Do not leave a high-risk run unattended unless local policy permits and risks are controlled. -
Wait for full stop and safe unlock
Never force the lid. If a power interruption occurs, follow the IFU for safe lid release. -
Unload carefully and verify separation
Remove tubes gently to avoid remixing layers. Check for hemolysis, tube cracks, or leakage. Proceed per SOP (aliquoting, transport to analyzer, or storage). -
Document the run if required
Record run parameters and any deviations, including alarms, delays, or unusual observations.
Typical settings and what they generally mean
Common programmable parameters include:
- Temperature setpoint: The target internal temperature; achieving and maintaining it depends on ambient conditions, rotor type, and speed. Some protocols require pre-cooling or a stabilization period.
- RCF or RPM: The “force” applied; RCF is more comparable across different rotor radii.
- Run time: Duration at target speed; some devices include ramp-up/ramp-down time in displayed run time, others do not (varies by manufacturer).
- Acceleration/deceleration (brake): Faster braking can resuspend pellets or disturb layers; a softer brake is often used for delicate separations, per SOP.
- Program memory: Stored protocols reduce error but can introduce risk if programs are mislabeled or overwritten.
If you are a trainee, treat these settings like medication dosing: follow the validated “order” (the SOP), and do not estimate.
Calibration and verification (what’s “relevant” in practice)
Users do not usually “calibrate” a centrifuge during routine operation, but facilities may perform verification checks such as:
- Speed verification using a tachometer (method and frequency vary by policy).
- Temperature verification using an independent probe (especially for critical workflows).
- Timer verification if the timer is used as a controlled parameter.
These checks are commonly organized by biomedical engineering or lab quality teams, not by individual clinicians. The need and frequency depend on accreditation expectations and risk assessment.
Temperature management tips (practical, non-brand-specific)
- Pre-cool when required by SOP; a cold rotor can be as important as a cold chamber.
- Minimize door-open time; repeated opening can drive condensation and temperature drift.
- Avoid overloading beyond rated capacity; dense loads can affect cooling performance.
- Recognize that high-speed runs generate heat; “setpoint equals sample temperature” is not always true (performance varies by manufacturer, rotor, and conditions).
Common model-to-model differences to watch for
- Some models prioritize RCF, others RPM; ensure you know which is displayed.
- Lid release mechanisms differ; emergency release procedures are not universal.
- Alarm codes are manufacturer-specific; keep the quick reference accessible.
- Some units log runs electronically; others rely on manual logbooks.
- Noise, vibration tolerance, and refrigeration behavior vary—especially between bench-top and floor-standing designs.
How do I keep the patient safe?
Even though the Refrigerated centrifuge is usually not patient-contacting, its output affects clinical decisions. Patient safety risks are often indirect: wrong sample, wrong processing, contamination, or delayed diagnostics. Safety is therefore both a biosafety issue and a quality-of-results issue.
Safety practices that protect patients through reliable lab results
Key practices include:
- Correct patient identification and labeling at collection and before centrifugation.
- Right tube, right handling: follow tube type requirements (additives, clot time where applicable, mixing technique) per SOP.
- Time and temperature control: manage pre-centrifugation delays and storage conditions according to validated workflows.
- Standardized centrifugation parameters: use protocol-defined RCF, time, temperature, and brake settings.
Small deviations can create large downstream effects in some assays. When in doubt, consult laboratory supervision rather than trying to “make the sample work.”
Biosafety: preventing exposure and contamination
Centrifugation can generate aerosols if a tube breaks or leaks. Controls typically include:
- Engineering controls: sealed rotors, aerosol-tight lids, safety cups, lid interlocks, imbalance detection.
- Administrative controls: SOPs for infectious specimens, restricted access, and training.
- PPE: gloves and eye protection as baseline; enhanced PPE for higher-risk specimens per policy.
- Work practice controls: allowing aerosols to settle before opening sealed accessories when required by SOP; opening rotors in appropriate containment if specified.
The correct combination depends on specimen risk category and facility biosafety level. Local policy should define when aerosol-tight accessories are mandatory.
Alarm handling and human factors (how errors happen)
Common human-factor failure modes include:
- Selecting the wrong stored program (similar names, legacy protocols)
- Confusing RPM with RCF
- Balancing by volume instead of mass
- Using the wrong adapter size “just this once”
- Rushing unloading and remixing layers
- Skipping documentation because the lab is busy
Risk controls that help:
- Clear labeling of rotors and adapters
- Color-coding or physical segregation of tube systems
- Program lockout features (if available) and standardized naming conventions
- Two-person verification for high-risk workflows (policy-dependent)
- “Stop-the-line” culture where staff can pause processing if something feels wrong
Labeling checks and traceability
Traceability is a patient safety tool. Practical steps include:
- Confirming sample ID matches requisition/electronic order before centrifugation
- Maintaining chain of custody for critical specimens (facility-dependent)
- Recording rotor and run parameters when required for quality investigations
- Using barcoding and Laboratory Information Systems (LIS) where available
If the facility uses decentralized centrifugation (for example, in emergency departments or satellite clinics), governance should clarify who owns SOP adherence and quality monitoring.
Incident reporting culture (general expectations)
When something goes wrong—tube breakage, repeated alarms, suspected temperature deviation, mislabeling—reporting should be:
- Non-punitive and timely
- Focused on learning and system improvement
- Documented in the facility reporting system per policy
For administrators, encourage reporting by ensuring follow-up actions are visible: staff are more likely to report if they see hazards addressed.
How do I interpret the output?
A Refrigerated centrifuge does not “diagnose” anything. Its “output” is primarily:
- A physically separated specimen (layers, pellets, clarified supernatant)
- Run parameters and device status (speed/RCF, time, temperature, alarms, logs)
Interpretation is therefore about sample quality and process confirmation, not clinical interpretation by itself.
Types of outputs/readings you may see
Depending on the model, the device may display:
- Set and actual temperature
- Set and actual RPM or RCF
- Remaining or elapsed time
- Acceleration/deceleration settings (sometimes as levels)
- Rotor identification (manual or automated; varies by manufacturer)
- Error codes and run history (varies)
Some facilities also use external verification tools during quality checks (for example, independent temperature probes), but that is typically a controlled process rather than a daily task.
How clinicians and lab teams typically interpret the separated specimen
Common observations include:
- Clear separation of serum/plasma from cellular components (where expected)
- Buffy coat visibility (white layer between plasma and red cells) in whole blood separations
- Pellet formation at the tube bottom (for cell pelleting workflows)
- Clarity of supernatant (turbidity may suggest incomplete separation, lipid content, or contamination—interpretation depends on context)
These observations support decisions like whether a sample is acceptable for analysis, needs re-centrifugation per SOP, or should be rejected and recollected based on policy.
Common pitfalls, limitations, and artifacts
- Incomplete separation: may occur with incorrect force/time, overfilled tubes, wrong rotor type, or brake settings that remix layers.
- Hemolysis: can be influenced by collection technique, transport, and centrifugation conditions; centrifugation is only one part of the chain.
- Temperature drift: actual sample temperature may differ from setpoint due to load and frictional heating; protocols may include mitigations (pre-cooling, defined delays).
- Tube incompatibility: tubes not rated for the applied force may deform or leak, compromising sample and creating biosafety hazards.
- Aerosol risk: opening a rotor after a tube failure without proper containment can expose staff and contaminate the environment.
Emphasize artifacts and the need for clinical correlation
Laboratory results derived from centrifuged samples still require clinical correlation. From an operations standpoint, the key principle is: if pre-analytical handling is suspect, the reliability of downstream results may be reduced. Follow facility policy on sample rejection criteria and repeat sampling.
What if something goes wrong?
When problems occur with a Refrigerated centrifuge, the response should prioritize staff safety, specimen integrity, and equipment protection. Do not troubleshoot in a way that bypasses safety interlocks or exposes staff to aerosols.
When to stop use immediately
Stop the run (or do not start) and remove the device from service if:
- There is unusual vibration, knocking, scraping, or burning smell
- The device shows repeated imbalance that is not resolved by proper balancing
- The lid lock behaves abnormally (won’t lock, won’t unlock appropriately)
- There is evidence of tube breakage, leakage, or internal spill
- Temperature control is clearly failing for a workflow that requires it (confirm per SOP)
- Any safety-critical alarm appears and you do not understand it
In many facilities, a “Do Not Use” tag and a rapid notification to biomedical engineering are standard.
A practical troubleshooting checklist (first-line)
Use local SOPs and the manufacturer’s guidance first. Common checks include:
- Power and restart: confirm power supply, breaker status, and that the unit completed its self-test.
- Lid issues: ensure nothing obstructs the lid; check gasket alignment; do not force.
- Imbalance errors: re-check symmetry, mass balance, tube seating, and correct adapters.
- Vibration: verify rotor is properly seated and tightened; check buckets swing freely; ensure the surface is level.
- Temperature not reaching setpoint: confirm ventilation clearance; reduce door-open frequency; allow pre-cooling time; check for overloaded chamber; confirm ambient conditions are within spec.
- Unexpected sample results: verify you used the correct RCF/time/brake and that the correct rotor radius assumptions were applied.
If the issue persists after safe, basic checks, stop and escalate.
When to escalate to biomedical engineering or the manufacturer
Escalate when:
- Error codes persist or are safety-related
- Refrigeration performance is abnormal (icing, persistent condensation, inability to stabilize)
- There is suspected mechanical damage (rotor, spindle, lid lock)
- The unit fails verification checks required by your quality system
- The problem recurs across multiple operators or runs
Biomedical engineering can coordinate vendor service, manage warranty/service contract processes, and document corrective actions (often within a CAPA process—Corrective and Preventive Action—depending on the facility quality framework).
Documentation and safety reporting expectations (general)
After an incident:
- Document what happened, including run parameters, rotor type, sample type, and alarms.
- Record actions taken: who was notified, decontamination performed, and whether samples were quarantined.
- Report in the facility incident reporting system per policy, especially for exposure, spills, or near-misses.
- Preserve evidence if needed (for example, damaged tubes or photos) according to local policy and privacy rules.
For administrators, ensure downtime procedures are clear: where samples go, how priorities are managed, and how turnaround time impacts are communicated clinically.
Infection control and cleaning of Refrigerated centrifuge
Cleaning and disinfection of a Refrigerated centrifuge should be treated as a routine safety process, not an afterthought. The key is to protect staff, prevent cross-contamination, and avoid damaging sensitive materials (rotors and chamber surfaces can be harmed by incompatible chemicals).
Cleaning vs disinfection vs sterilization (general)
- Cleaning removes visible soil and organic material (usually with detergent and water). Cleaning is often required before disinfection can be effective.
- Disinfection reduces microorganisms on surfaces using chemical agents. The required level (low/intermediate/high) depends on the risk assessment and local policy.
- Sterilization eliminates all microbial life, typically using heat, gas, or radiation. Centrifuge chambers and rotors are not usually sterilized as whole units in routine clinical practice; certain removable components may have specific reprocessing guidance. Always follow the IFU.
The correct approach depends on specimen risk and the facility infection prevention policy.
High-touch points and high-risk areas
Common high-touch points:
- Lid handle and outer lid surface
- Control panel/buttons/touchscreen
- Emergency stop and power switch (if present)
- Front edges and side handles used to move the unit
High-risk internal areas:
- Chamber bowl and drain areas (if present)
- Rotor surfaces, buckets, adapters, and seals
- Under rotor and around the spindle (where spills can be missed)
Example cleaning workflow (non-brand-specific)
Use facility-approved products and follow the manufacturer IFU. A general workflow might include:
-
Preparation
Put on appropriate PPE. If dealing with a suspected infectious spill, follow biosafety spill policy. -
Power safety
Stop the unit and disconnect power if required by SOP and safe to do so. Allow the rotor to fully stop and unlock normally. -
Remove accessories
Remove rotor/buckets/adapters carefully. Inspect for spills, cracks, and residue. -
Clean (detergent step)
Wipe internal chamber and removable components with a compatible detergent solution. Avoid abrasive pads that can scratch surfaces. -
Disinfect
Apply the facility-approved disinfectant at the correct contact time. Ensure you reach seams, grooves, and gasket areas. -
Rinse/wipe as required
Some disinfectants require rinsing to prevent corrosion. Follow IFU and infection prevention guidance. -
Dry thoroughly
Moisture can contribute to corrosion, odors, and microbial growth. Allow adequate drying time before closing the lid. -
Reassemble and document
Reinstall rotor and accessories correctly. Document cleaning per policy (especially after spills or high-risk use).
Managing spills and tube breakage (biosafety-focused)
If a tube breaks or leaks:
- Treat it as a potential aerosol event.
- Keep the lid closed for the time specified by local policy (some protocols include a settling period).
- Use PPE appropriate to the hazard level.
- Open and clean using containment practices required by your biosafety team (this may include moving sealed components to a biological safety cabinet if specified and feasible).
- Quarantine the device if needed until decontamination and inspection are complete.
Important note on chemical compatibility
Do not assume all disinfectants are safe for all rotors and chamber materials. Some agents can accelerate corrosion or degrade seals. Use only chemicals approved by the manufacturer IFU and facility policy. When uncertain, consult biomedical engineering and infection prevention.
Medical Device Companies & OEMs
Manufacturer vs. OEM (Original Equipment Manufacturer)
A manufacturer is the company that markets and supports the final product under its name, including labeling, documentation, warranties, and regulatory responsibilities (as applicable in a given jurisdiction). An OEM (Original Equipment Manufacturer) is an entity that makes a component or an entire device that may be branded and sold by another company.
In practice, one Refrigerated centrifuge model may be:
- Designed, manufactured, and supported by one company, or
- Manufactured by an OEM and sold under multiple brand names, or
- Assembled from globally sourced components (compressors, electronics, rotors), with final branding and support handled by a separate entity
These relationships are common across medical equipment and laboratory instrumentation.
How OEM relationships impact quality, support, and service
For hospitals and labs, OEM structures can influence:
- Parts availability and lead times
- Service network maturity and training depth for field engineers
- Software/firmware updates and cybersecurity patch practices (where applicable)
- Consistency of accessories (rotors, buckets, adapters) across product lines
- Warranty clarity: who is responsible for what, and under which conditions
A practical procurement question is not only “Who sells it?” but also “Who services it locally, and where do parts come from?” This becomes especially important in regions with import restrictions, long shipping times, or limited authorized service coverage.
Top 5 World Best Medical Device Companies / Manufacturers
The list below is example industry leaders (not a ranking) commonly associated with laboratory centrifugation and broader scientific/clinical device portfolios. Product availability, local support quality, and regulatory status vary by manufacturer and country.
Thermo Fisher Scientific
Thermo Fisher Scientific is widely recognized for laboratory and clinical laboratory infrastructure, with product lines that often include centrifuges as well as reagents and analytical instruments. In many regions, it operates both as a manufacturer and through distributor channels, which can simplify bundled procurement for some buyers. Service coverage and parts logistics can be strong in major urban markets, while remote coverage varies by country and contract structure. Specific Refrigerated centrifuge model features and certifications vary by region and intended use.
Eppendorf
Eppendorf is well known in many laboratories for benchtop instruments and consumables, including centrifugation solutions used in clinical-adjacent and research workflows. Its reputation is often associated with usability and laboratory ergonomics, though exact performance characteristics depend on model and rotor configuration. Global availability is typically supported through a mix of direct offices and authorized distributors. Accessory ecosystems (rotors, adapters) are an important part of the purchasing decision and should be confirmed during procurement.
Beckman Coulter (Danaher)
Beckman Coulter is associated with clinical diagnostics and life science instrumentation, with centrifugation historically forming part of many laboratory workflows. In some settings, buyers value integrated service structures and the ability to align centrifugation platforms with other laboratory systems, though this is facility-dependent. As with all manufacturers, local service experience can differ between countries and even between cities. Confirm intended-use labeling and support terms for the specific Refrigerated centrifuge model being considered.
Andreas Hettich GmbH & Co. KG (Hettich)
Hettich is known in many markets for centrifugation systems used in clinical laboratories, blood banks, and research environments. Buyers often evaluate Hettich for rotor variety and workflow fit, especially where multiple tube formats must be supported. Distribution and service are frequently handled via regional partners, making due diligence on local support essential. Exact product configurations and options vary by manufacturer and region.
Sigma Laborzentrifugen (Sigma)
Sigma is recognized in several markets for laboratory centrifuges, including refrigerated configurations used in clinical and research-adjacent workflows. As with other centrifuge-focused manufacturers, the practical purchasing factors often include rotor availability, aerosol containment options, and serviceability. Global reach may depend significantly on authorized distributors and service partners in each country. Confirm spare-parts pathways and preventive maintenance expectations during procurement planning.
Vendors, Suppliers, and Distributors
Role differences between vendor, supplier, and distributor
These terms are sometimes used interchangeably, but they can mean different things operationally:
- Vendor: the entity you buy from (could be the manufacturer, a reseller, or a tender-awarded contractor).
- Supplier: the organization that provides goods/services to your facility; may include consumables, accessories, and service.
- Distributor: a company that stocks, imports, and resells products—often providing logistics, local invoicing, and sometimes first-line technical support.
For a Refrigerated centrifuge purchase, a hospital might contract with a distributor for the device, but rely on the manufacturer (or a certified service partner) for warranty repairs and specialized training. Clarifying roles up front reduces delays during downtime events.
Top 5 World Best Vendors / Suppliers / Distributors
The list below is example global distributors (not a ranking) that are commonly associated with laboratory and healthcare supply chains. Actual availability of Refrigerated centrifuge models, service arrangements, and country coverage vary substantially.
Avantor (VWR)
Avantor (including the VWR brand in many markets) is widely known for supplying laboratory consumables and equipment through broad catalogs. In some regions it supports institutional buyers such as hospitals, universities, and reference laboratories with consolidated purchasing and logistics services. Whether it supplies specific Refrigerated centrifuge brands depends on local authorizations and contracts. Service coordination may be provided directly or through manufacturer partnerships, depending on geography.
Fisher Scientific (Thermo Fisher channel)
Fisher Scientific is commonly recognized as a major channel for laboratory products in many markets, with procurement-friendly catalog systems and institutional account management. For some buyers, bundled purchasing of reagents, consumables, and equipment through a single channel can simplify operations. Availability of Refrigerated centrifuge units and accessories depends on local inventory strategies and manufacturer distribution agreements. Service support pathways should be clarified in writing, especially for critical laboratory uptime needs.
DKSH
DKSH is known in parts of Asia (and selected other regions) for market expansion services that can include distribution, logistics, and after-sales support coordination. In practice, hospitals in emerging markets may encounter DKSH as a channel partner for global medical equipment brands. Coverage is often stronger in major cities than in remote areas, so service planning should include response-time expectations. As always, confirm whether the distributor is authorized for the specific brand and model.
Henry Schein
Henry Schein is recognized in multiple countries for healthcare distribution, historically strong in dental and clinic-based supply chains, with some medical equipment offerings depending on region. Where it participates in hospital procurement, it may function as a vendor for selected equipment categories and a broad supplier for consumables. Availability of Refrigerated centrifuge lines is region-dependent and not universal. Buyers should confirm technical service arrangements rather than assuming in-house service capacity.
Bunzl (healthcare and safety supply channels)
Bunzl operates internationally in distribution across several sectors, including healthcare and safety supplies in certain markets. For hospitals, it may be more visible as a supplier of consumables, infection prevention products, and logistics services than as a primary source of specialized laboratory instruments. Where it is involved in equipment supply, the role may be as a vendor coordinating with manufacturers and service partners. As with all distributors, confirm authorization status, warranty handling, and spare-parts pathways.
Global Market Snapshot by Country
India
Demand for Refrigerated centrifuge systems in India is closely tied to growth in private diagnostic chains, expanding hospital laboratory capacity, and academic research output. Many facilities rely on a mix of imported brands and locally assembled options, with procurement influenced by tender processes and service reach. Service ecosystems are generally stronger in major cities, while rural access can depend on distributor networks and biomedical engineering capacity.
China
China has substantial domestic manufacturing capacity across laboratory and clinical device categories, alongside ongoing demand for imported Refrigerated centrifuge models in certain segments. Public hospital procurement processes, local standards, and price competitiveness shape buying decisions. Service coverage is often robust in urban centers, while remote regions may experience longer repair cycles depending on parts logistics.
United States
The United States has a mature installed base of Refrigerated centrifuge units across hospital labs, reference labs, and academic centers, with strong emphasis on standardized workflows and documented quality practices. Buyers often prioritize uptime, service contract performance, and accessory standardization across multiple sites. Rural access is generally supported through regional service networks, though response times can vary by contract and geography.
Indonesia
Indonesia’s archipelagic geography makes logistics and service coverage a key factor for Refrigerated centrifuge procurement, especially outside major urban areas. Import dependence is common for higher-end laboratory instruments, and distributor strength can strongly influence downtime risk. Urban private hospitals and reference labs typically have better access to training and service than remote facilities.
Pakistan
In Pakistan, demand is driven by expanding private laboratories, tertiary hospitals, and public health needs, with many Refrigerated centrifuge purchases routed through local distributors. Import processes and currency variability can affect lead times for equipment and spare parts. Service capability is often concentrated in major cities, making preventive maintenance planning important for facilities in smaller regions.
Nigeria
Nigeria’s market for Refrigerated centrifuge systems is influenced by growth in private diagnostics, teaching hospitals, and public health laboratory strengthening initiatives. Import dependence and power reliability concerns shape purchasing priorities, often pushing buyers toward robust designs and clear service support. Service and parts availability may be uneven outside major urban areas, increasing the importance of local technical partnerships.
Brazil
Brazil’s demand spans public health networks, private hospital groups, and large reference laboratories, with procurement influenced by regulatory requirements and import considerations. Some buyers prioritize vendor-managed service and local parts stocking due to the operational impact of downtime. Access to service is typically better in major metropolitan regions than in remote areas.
Bangladesh
Bangladesh continues to expand diagnostic services in urban centers, with Refrigerated centrifuge demand linked to hospital growth, private lab investment, and public health initiatives. Import dependence is common, and procurement teams often weigh total cost of ownership alongside upfront pricing. Service ecosystems may be concentrated in major cities, making training and preventive maintenance planning essential for regional facilities.
Russia
Russia’s Refrigerated centrifuge market is shaped by public healthcare infrastructure, academic research, and evolving import conditions that can influence brand availability and spare-parts access. Facilities may place increased emphasis on local service capability and parts substitution strategies within policy limits. Urban centers generally have stronger service coverage than more remote regions.
Mexico
Mexico’s demand reflects a mix of public sector procurement and private laboratory networks, with supply chains benefiting from proximity to North American manufacturing and distribution in some cases. Import processes and tender structures can influence model selection and accessory standardization. Service availability is typically better in large cities, with regional coverage depending on distributor footprints.
Ethiopia
Ethiopia’s Refrigerated centrifuge needs are tied to expanding hospital capacity, laboratory strengthening programs, and public health testing infrastructure. Import dependence is common, and long-term maintenance planning is critical where local service resources are limited. Urban tertiary centers may have stronger biomedical engineering support than rural facilities, affecting equipment selection and resilience requirements.
Japan
Japan represents a highly standardized and quality-focused market, with strong expectations around device reliability, documentation, and service responsiveness. Refrigerated centrifuge procurement often emphasizes long lifecycle management, accessory compatibility, and predictable performance in controlled laboratory environments. Rural access is generally supported by established service networks, though specific support models vary by manufacturer.
Philippines
In the Philippines, demand is driven by private hospital expansion, reference laboratories, and public health capacity needs, with many Refrigerated centrifuge units sourced through distributors. Geographic dispersion across islands makes logistics and service coverage important, particularly for parts and rotor availability. Major urban centers typically have better access to trained service personnel than remote areas.
Egypt
Egypt’s market includes large public hospital systems and a growing private sector, both of which require reliable laboratory throughput and biosafety controls. Refrigerated centrifuge procurement can be influenced by tendering, import processes, and the availability of authorized service partners. Service access is often stronger in major cities, while regional facilities may experience longer repair timelines.
Democratic Republic of the Congo
In the Democratic Republic of the Congo, Refrigerated centrifuge demand is closely linked to essential laboratory services, outbreak response readiness, and donor-supported health programs in some contexts. Import dependence and infrastructure variability (power, transport, service workforce) strongly shape what models are practical. Facilities often prioritize ruggedness, straightforward maintenance, and locally feasible training approaches.
Vietnam
Vietnam’s expanding hospital sector and growing diagnostic laboratory capacity are major drivers for Refrigerated centrifuge adoption, alongside academic and biotech activity. Many facilities procure through local distributors who provide import handling and first-line support, making distributor capability a major differentiator. Urban centers typically have better access to service and training than provincial sites.
Iran
Iran’s market is shaped by a combination of local production capacity in selected categories and variable access to imported medical equipment depending on trade and regulatory conditions. Refrigerated centrifuge procurement often emphasizes serviceability and parts pathways that remain reliable over time. Access to advanced service resources may differ between major cities and smaller regions.
Turkey
Turkey has a large healthcare delivery system and a significant medical manufacturing and distribution presence, supporting a diverse market for Refrigerated centrifuge systems. Procurement is influenced by public tenders, private hospital investment, and regional export/import dynamics. Service ecosystems are often strong in urban areas, and many facilities prioritize clear maintenance and parts support arrangements.
Germany
Germany is a mature market with strong engineering and quality expectations across laboratory and hospital equipment. Refrigerated centrifuge purchasing decisions often emphasize compliance documentation, preventive maintenance discipline, and long-term lifecycle planning. Service availability is typically robust, and accessory standardization across networks is common in larger organizations.
Thailand
Thailand’s demand is supported by public health infrastructure, private hospital growth, and laboratory services aligned with medical tourism in some areas. Refrigerated centrifuge procurement commonly balances performance needs with service coverage and training support, especially for multi-site hospital groups. Urban centers generally have stronger distributor and service presence than rural regions.
Key Takeaways and Practical Checklist for Refrigerated centrifuge
- Treat the Refrigerated centrifuge as a safety-critical step in the diagnostic chain, not “just a lab tool.”
- Use only validated SOP parameters for the specific sample type, tube type, and rotor configuration.
- Prefer RCF-based protocols when possible, because RPM does not translate across different rotor radii.
- Verify the rotor, buckets, and adapters are compatible and rated for the planned force and temperature.
- Inspect rotors routinely for corrosion, cracks, and damage, and follow the manufacturer’s retirement criteria.
- Balance samples by mass and symmetry every time, even if the device has imbalance detection.
- Do not run damaged, poorly capped, or incompatible tubes that can leak or break under force.
- Pre-cool the unit or rotor when required; refrigeration setpoint does not guarantee sample temperature.
- Minimize lid-open time to reduce temperature drift and condensation inside the chamber.
- Choose brake/deceleration settings that match the protocol to avoid remixing layers after separation.
- Keep aerosol-tight accessories available and use them when specimen risk or policy requires.
- Stay nearby during startup to identify immediate imbalance or abnormal vibration early.
- Never bypass lid locks or interlocks, and never force the lid open.
- Stop use if you hear unusual noise, feel excessive vibration, or see repeated error codes.
- Treat tube breakage as a potential aerosol event and follow spill containment procedures.
- Clean and disinfect after spills promptly, using only chemicals compatible with the device materials.
- Document cleaning, runs, and anomalies according to your laboratory quality system.
- Standardize tube types and rotor systems across sites to simplify training and inventory control.
- Keep quick-reference SOPs and alarm code guides accessible at point of use.
- Use clear naming and access controls for stored programs to reduce wrong-program selection.
- Build competency training that includes balancing, rotor selection, biosafety, and troubleshooting basics.
- Separate responsibilities: lab owns SOPs, biomedical engineering owns maintenance coordination, procurement owns lifecycle planning.
- Plan for downtime with backup centrifugation capacity or sample rerouting procedures.
- Confirm service coverage, response times, and spare parts pathways before purchase or renewal.
- Include accessory and rotor costs in total cost of ownership, not just the base device price.
- Ensure the installation site meets ventilation, clearance, power, and ambient condition requirements.
- Consider ergonomics and safe lifting practices for heavy rotors to reduce staff injury risk.
- Use a rotor log when required, especially for high-speed or high-criticality applications.
- Verify temperature and speed performance per your facility’s commissioning and periodic checks policy.
- Quarantine suspect samples when centrifugation deviations occur and follow lab guidance on acceptability.
- Promote a non-punitive incident reporting culture for near-misses and equipment hazards.
- Align infection prevention policy with centrifugation realities, especially for aerosol-generating risks.
- Train staff to recognize the difference between device display setpoint and actual sample conditions.
- Keep consumables (tubes, adapters, seals) stocked to avoid unsafe substitutions during busy shifts.
- Use checklists at the point of use to reduce cognitive load during high-throughput periods.
- Audit compliance periodically and use findings to improve SOP clarity and workflow design.
- Review procurement contracts for clarity on who provides training, validation support, and software updates.
- Ensure emergency lid release procedures are known, documented, and used only when safe and indicated.
- Treat repeated minor alarms as early warnings and escalate before they become major failures.
- Reassess centrifuge placement if vibration, heat buildup, or workflow congestion becomes recurring problems.
- Confirm local regulatory and accreditation expectations for documentation and verification activities.
- When uncertain, pause and consult laboratory leadership rather than improvising centrifugation parameters.
If you are looking for contributions and suggestion for this content please drop an email to contact@myhospitalnow.com
Find Trusted Cardiac Hospitals
Compare heart hospitals by city and services — all in one place.
Explore Hospitals